EasyManua.ls Logo

Rohde & Schwarz ZNB40 - Calculate<Ch>:Transform:vnetworks:gloop:embedding<Group>[:State]

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
966User Manual 1173.9163.02 ─ 54
Parameters for setting and query:
<CircuitModel> SL | SC
Possible circuit models (character data); see Table 8-8.
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:R? SC; R? SL
Query the default resistances for ground loop deembedding. The
response is 10000000; 0.
CALC:TRAN:VNET:GLO:DEEM:PAR:R SC, 2.2E+3
Increase the resistance for the Shunt C model to 2.2 kΩ.
Manual operation: See "Network" on page 610
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
TNDefinition <CircuitModel>
Selects the circuit model for ground loop deembedding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<CircuitModel> FIMPort | SL | SC | SG
Possible circuit models (character data); see Circuit models for
ground loop port embedding/deembedding.
Example: See CALCulate<Ch>:TRANsform:VNETworks:GLOop:
DEEMbedding<group>[:STATe] on page 963
Manual operation: See "Network" on page 610
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>[:STATe]
<Boolean>
Enables or disables the embedding function for ground loops. It is allowed to change
the circuit model and its parameters while embedding is enabled.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
SCPI Command Reference

Table of Contents

Related product manuals