EasyManua.ls Logo

Rohde & Schwarz ZNB40 - Calculate<Ch>:Transform:vnetworks:sended:embedding<Phypt>:Parameters: G<Cmp; Calculate<Ch>:Transform:vnetworks:sended:embedding<Phypt>:Parameters: L<Cmp; Calculate<Ch>:Transform:vnetworks:sended:embedding<Phypt>:Parameters: R<Cmp; Calculate<Ch>:Transform:vnetworks:sended:embedding<Phypt>:Tndefinition

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
986User Manual 1173.9163.02 ─ 54
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Parameters:
<Boolean> ON - embedding active
OFF - embedding inactive
*RST: OFF
Example:
CALC:TRAN:VNET:SEND:EMB:TND CSL
Select the Serial C, shunt L circuit model for embedding.
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3; :
CALC:TRAN:VNET:SEND:EMB ON
Increase the resistance R2 for the Serial C, shunt L circuit model
to 2.2 kΩ and enable embedding.
Manual operation: See "Active" on page 574
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA <Interchange>, <arg1>
Defines a dembedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition to select the adequate circuit model before executing this com-
mand.
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt> to load circuit
data from a Touchstone file located at the R&S ZNB/ZNBT's file system.
Suffix:
<Ch>
.
Channel number
<PhyPt> Physical port number
Setting parameters:
<Interchange> FPORts | IPORts | SGATes
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<arg1> <block_data>
Content of a Touchstone file (*.s2p or *.s4p) in IEEE488.2
Block Data Format.
Usage: Setting only
SCPI Command Reference

Table of Contents

Related product manuals