Rohde & Schwarz ZNB40 - Calculate<Chn>:Transform:complex; Calculate<Chn>:Transform:impedance:rnormal; Calculate<Chn>:Transform:time[:Type]

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
931User Manual 1173.9163.02 ─ 54
Parameters:
<SidebandSupp> Sideband suppression
Range: 10 dB to 120 dB
Increment: 10 dB
*RST: 32 dB
Default unit: dB
Example:
*RST; :CALC:TRAN:TIME:WIND DCH
Reset the instrument and select a Dolph-Chebyshev window for
filtering the data in the frequency domain.
CALC:TRAN:TIME:DCH 25
Set the sideband suppression to 25 dB.
Manual operation: See "Side Lobe Level" on page 320
CALCulate<Chn>:TRANsform:TIME:LPASs <Algorithm>
Calculates the harmonic grid for low pass time domain transforms according to one of
the three alternative algorithms.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Algorithm> KFSTop | KDFRequency | KSDFrequency
KFSTop - keep stop frequency and number of points
KDFRequency - keep frequency gap and number of points
KSDfrequency - keep stop frequency and approximate fre-
quency gap
Example: See CALCulate<Chn>:TRANsform:TIME[:TYPE]
Manual operation: See "Set Harmonic Grid and Keep" on page 322
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam <DCValue>
Defines the DC value for low pass transforms. The command is enabled only if the
sweep points are on a harmonic grid (to be set explicitly or using CALCulate<Chn>:
TRANsform:TIME:LPASs).
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DCValue> DC value of the measured quantity
Range: Depending on the measured quantity (-1 to +1 for
S-parameters)
*RST: 0
SCPI Command Reference