Command Reference
R&S
®
ZNB/ZNBT
846User Manual 1173.9163.02 ─ 54
CALCulate:FMODel:DELT<Ph_pt>:SWEep:CONTrol:STEP?
For the Delta-L measurement, the firmware can either use a default frequency step
size, or the active channel setting (see CALCulate:FMODel:DELT<Ph_pt>:SWEep:
CONTrol).
This query returns the resulting frequency step size.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Query only
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Sweep Control" on page 616
CALCulate:FMODel:DELT<Ph_pt>:SWEep:FREQuency:MAXimum
<Frequency>
CALCulate:FMODel:DELT<Ph_pt>:SWEep:FREQuency:MINimum <Frequency>
Defines the sweep range of the Delta-L measurement channel. The entered MAXimum
and MINimum values are set as [SENSe<Ch>:]FREQuency:STOP and
[SENSe<Ch>:]FREQuency:STARt, respectively.
Furthermore, the VNA collects data for the LPI computation a the MAXimum and MINi-
mum frequencies (in addition to CALCulate:FMODel:DELT<Ph_pt>:
FREQuencies:CURRent?).
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Frequency> Frequency values
The default MINimum frequency is the standard Delta-L start fre-
quency of 10 MHz. The default MAXimum frequency is either 20
GHz or the instrument's maximum frequency, whichever is
smaller.
Default unit: Hz
Options: R&S ZNB/ZNBT-K231
Manual operation: See "Sweep Frequency" on page 616
CALCulate:FMODel:DELT<Ph_pt>:TCONfig <Boolean>
Defines how the analyzer firmware sets up the diagram area for the Delta-L measure-
ment.
Suffix:
<Ph_pt>
.
This suffix is ignored
Parameters:
<Boolean>