EasyManua.ls Logo

Rohde & Schwarz ZNB40 - Calculate<Chn>:Statistics[:State]; Calculate<Chn>:Statistics[:State]:Area

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
919User Manual 1173.9163.02 ─ 54
Parameters:
<Boolean> ON | OFF - Statistical info field on or off.
*RST: OFF
Example:
*RST; :CALC:STAT:MMPT ON
Reset the instrument, hiding all statistical results. Display the
"Min/Max/Peak-Peak" results.
CALC:STAT:MSTD ON
Display the "Mean/Std Dev" results in addition.
CALC:STAT:RMS ON
Display the "RMS" results in addition.
CALC:STAT:EPD ON
Display the "Phase/El Length" results in addition.
CALC:STAT:SFL ON
Display the "Flatness/Gain/Slope" results in addition.
CALC:STAT:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
...
CALC:STAT OFF
Hide all results.
Manual operation:
See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 326
CALCulate<Chn>:STATistics[:STATe]:AREA <HorizontalPos>, <VerticalPos>
Moves the statistics info field for the active trace <Chn> to one of nine predefined posi-
tions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:STATistics[:STATe]
Manual operation: See "Min/Max/Peak-Peak, Mean/Std Dev/RMS" on page 326
CALCulate<Chn>:STATistics:DOMain:USER <EvalRange>
Selects one out of 10 evaluation ranges to be configured with the CALCulate<Chn>:
STATistics:DOMain:USER:SHOW, CALCulate<Chn>:STATistics:DOMain:
USER:STARt, and CALCulate<Chn>:STATistics:DOMain:USER:STOP com-
mands.
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI Command Reference

Table of Contents

Related product manuals