Command Reference
R&S
®
ZNB/ZNBT
961User Manual 1173.9163.02 ─ 54
FPORts (or omitted)
Standard port sequence (network port 1 towards VNA, network
port 2 towards DUT)
IPORts | SGATes
Inverted port sequence (network port 2 towards VNA, network
port 1 towards DUT)
<SParamTrcs> <block_data>
Content of a two-port Touchstone file (*.s2p) in IEEE488.2
Block Data Format.
Usage: Setting only
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:G<Cmp> <CircuitModel>[, <Conductance>]
Specifies the conductance value G in the "Shunt L, Shunt C" lumped element model
for differential match embedding.
In the query form, the <Conductance> parameter must be omitted.
Suffix:
<Ch>
.
Channel number
<LogPt> Logical port number (balanced port)
<Cmp> Must be omitted or set to 1.
Parameters:
<CircuitModel> SHLC
Currently only the "Shunt L, Shunt C" lumped element model is
supported
<Conductance> Range: -10 MS to 10 MS.
Increment: 1 μS (1E-6 F)
*RST: 0 S
Default unit: S(iemens)
Manual operation: See "Network" on page 613
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:L<Cmp> <CircuitModel>[, <Inductance>]
Specifies the inductance value L in the "Shunt L, Shunt C" lumped element model for
differential match embedding.
In the query form, the <Inductance> parameter must be omitted.