Rohde & Schwarz ZNB40 - Calculate<Ch>:Transform:vnetworks:gloop:deembedding<Group>:Parameters:C; Calculate<Ch>:Transform:vnetworks:gloop:deembedding<Group>:Parameters:g; Calculate<Ch>:Transform:vnetworks:gloop:deembedding<Group>:Parameters:l; Calculate<Ch>:Transform:vnetworks:gloop:deembedding<Group>:Parameters:r

Rohde & Schwarz ZNB40
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
963User Manual 1173.9163.02 ─ 54
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>[:STATe]
<Boolean>
Enables or disables the deembedding function for ground loops. It is allowed to change
the circuit model and its parameters while deembedding is enabled.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Boolean> ON - Deembedding active
OFF - Deembedding inactive
*RST: OFF
Example:
CALC:TRAN:VNET:GLO:DEEM:TND SL
Select the Shunt L circuit model for deembedding.
CALC:TRAN:VNET:GLO:DEEM:PAR:R SL, 2.2E+3; :
CALC:TRAN:VNET:GLO:DEEM ON
Increase the resistance for the Shunt L circuit model to 2.2 kΩ
and enable deembedding.
Manual operation: See "Active" on page 578
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C <CircuitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:C? <CircuitModel>
Specifies the capacitance value C in the different circuit models for ground loop
embedding.
Suffix:
<Ch>
.
Channel number
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Capacitance> Capacitance C for ground loop deembedding.
Range: -1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST: 1 pF (1E-12 F)
Default unit: F