- 17 -
CLEANING AND DISINFECTING
Cleaning and disinfecting are two separate processes. Cleaning MUST precede disinfection.
Note: Product MUST be cleaned, disinfected, and checked for proper functioning between use by different individuals.
CAUTION: DO NOT use abrasives such as steel wool or scouring pads on components of AGILITY.
DO NOT use cleaning products containing petroleum or organic solvents [i.e. acetone, toluene,
Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids, adhesive removers] on components of AGILITY.
WARNING: If instructed to use bleach, follow safety guidelines on bleach container.
AGILITY Component To Clean: To Disinfect:
Back shell, back shell hardware
and back cane hardware:
Hand wash in warm (40°C) water with soap and a
sponge. Rinse with clean water. Wipe dry with a
clean cloth and then allow to air dry completely.
Wipe gently with a cloth dampened with a
household disinfectant. Rinse with clean water.
Wipe dry with a clean cloth and then allow to air
dry completely.
Cover:
Notes: Remove the foam and
air components from the cover
prior to cleaning and disinfecting.
Secure the hook and loop fastener
strip on the handle at the wide
end of the cover.
Turn the cover inside out.
Zip the cover closed to avoid
snags.
Machine wash in warm (40°C) water with mild
detergent. Tumble dry low.
Hand wash in warm (40°C) water using 1 part
household liquid bleach per 9 parts water. Rinse
thoroughly with clean water. The cover may also
be machine washed in hot (60°C) water to
disinfect. Tumble dry low.
Foam pad, foam insert and
lumbar support pad:
Note: Remove the foam
components from the cover prior
to cleaning.
Note: Disinfectants are NOT effective on porous
surfaces such as foam. Replace soiled or
contaminated foam components with authorized
ROHO foam components only. Replacement foam
components may be purchased separately. Soiled
or contaminated foam components MUST NOT be
used by different individuals.
Wipe lightly with a damp cloth. DO NOT use soap.
DO NOT submerge in water. Wipe off with a clean
cloth. Completely air dry before replacing in cover.
Air insert:
Notes: Remove the air
insert from the cover and foam
pad prior to cleaning and
disinfecting.
Make sure the valve is closed.
DO NOT allow water or cleaning
solution to enter the air insert.
Place in large sink or basin with warm (40°C)
water. ROHO, Inc. recommends using a mild
hand-washing or dish-washing liquid, laundry
detergent, or other multipurpose detergents used
for general cleaning. Use a soft, plastic bristle
brush, sponge, or washcloth to gently scrub all
cushion surfaces. Rinse with clean water. Air dry.
DO NOT machine wash or dry. DO NOT expose to
direct sunlight.
Hand wash in warm (40°C) water using 1 part
household liquid bleach per 9 parts water. Rinse
thoroughly with clean water. Air dry. DO NOT
machine wash or dry. DO NOT autoclave.
= Wipe with damp cloth
1:9
+
1:9
or
+