MODEL 1054B pH/ORP SECTION 4.0
CONFIGURATION
TABLE 4-1. Configuration Work Sheet (pH)
Use this work sheet to assist in the configuration of the analyzer. Date: ____________________
RANGE FACTORY SET USER SET
A. Alarm 1 Setup (AL1)
1. Alarm Configuration (On/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Lo _________
3. Hysteresis (HYS) 0-2.0 pH 0 pH _________
4. Delay Time On (don) 0-255 sec. 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec. 000 Seconds _________
B. Alarm 2 Setup (AL2)
1. Alarm Configuration (On/FLt/OFF) On _________
2. High or Low (H-L) (Hi/Lo) Hi _________
3. Hysteresis (HYS) 0-2.0 pH 0 pH _________
4. Delay Time On (don) 0-255 sec 000 Seconds _________
5. Delay Time Off (doF) 0-255 sec 000 Seconds _________
C. Interval Timer (
IInntt
)
1. Active Status (tOn) (oFF/on) oFF _________
2. Interval Time (int) Minimum 10 Minutes 10 Seconds _________
3. Count (cnt) 1 to 60 5 _________
4. On Time (ont) 0 to 299 sec 1 Second _________
5. Off Time (OFt) 0 to 299 sec 1 Second _________
6. Duration (dur) 0 to 299 sec 2 Seconds _________
D. Temperature Setup (t-C)
1. Display Temperature (d-t) (
o
C/
o
F) *C _________
2. Automatic Temperature
Compensation (Atc) (on/oFF) on _________
a. Manual Temp. Value -10°C to 125°C _________
E. Current Output Setup (OUt)
1. mA Output (CUr) (020/420) 420 _________
2. Display Current Output (d-O) (Pct/doc) doc _________
3. Dampen Current Output (dPn) 0-255 sec. 0.0 Seconds _________
F. Electrode Diagnostics Setup (InP) (on/oFF) oFF
1. Temp compensated impedance (rES)
2.
Impedance increase before calibration
(CAL) 20-200% 20% _________
3. Cracked glass impedance low limit (SHt) 5-600 megohms 20 megohms _________
4. Aged electrode impedance high limit (OId)
200-1999 megohms
1000 megohms _________
G. Default Setup (dFt)
1. Relay 1 Default (rL1) (non/oFF/on) non _________
2. Relay 2 Default (rL2) (non/oFF/on) non _________
3. Current Output Default (Cur) (non/cur) non _________
H. Automatic Buffer Mode (AbF)
1. Configuration (on/off) on _________
I. Keyboard Security Setup (COd)
1. Keyboard Security Required 001-999
_
_________
2. Keyboard Security Not Required 000 000 _________
Alarm Setpoints
1. Alarm 1 (SP 1) 0-14 pH 0.00 pH _________
2. Alarm 2 (SP 2) 0-14 pH 14.00 pH _________
Current Output
1. Zero (0 or 4 mA) (LO) 0-14 pH 0.00 pH _________
2. F.S. (20 mA) (HI) 0-14 pH 14.00 pH _________