EasyManua.ls Logo

Rosemount 1054B - 4.10 Alarm Setpoint Adjustment

Rosemount 1054B
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
MODEL 1054B pH/ORP SECTION 4.0
CONFIGURATION
4.10 ALARM SETPOINT. The alarm setpoints should
be adjusted after completing the configuration proce-
dure as outlined in Sections 4.2 to 4.9.
1. Press the PV key to ensure that the analyzer is
not in Set Mode.
2. Press the ALARM 1 or ALARM 2 key. SP1 or SP2
will show briefly, followed by the Alarm 1 or
Alarm 2 setpoint.
NOTE
If the alarm is set to OFF or FAULT
(Alarm 2 only), the analyzer will display
OFF or FLt respectively (Refer to Section
4.2.1, Alarm Setup).
NOTE
Alarm logic may be changed from normal-
ly open (N.O.) to normally closed (N.C.) by
cutting circuits on the power supply PCB
(W-5, W-7, W-9) and adding jumpers (W-
4, W-6, W-8).
3. SELECT to adjust the value. The display will
acknowledge briefly with AdJ followed by the
Numeric Display with digit flashing.
4. SCROLL (
é
) and SHIFT (
ç
) to display the
desired value.
5. ENTER value into memory.
6. Repeat Steps 2 to 5 for the second setpoint.
ZERO
ALARM
1
F.S .
ALARM
2
ACCESS
é
AdJ
SP1/2
ç
SELECT
ç
SELECT
ENTER
Press
Once
Press
Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press
Once
Numeric
Display
of
Setpoint
FIGURE 4-4. Alarm Setpoint

Related product manuals