EasyManua.ls Logo

SA Instruments 1030 - microCapStar Capnograph

Default Icon
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Capnograph and Ventilator
Revision C, August 2013
14-3
The diagram above shows connections between the Control Unit and the Valve
Assembly as well as connections to components of an anesthesia setup. The ventilator
should be located near a waveguide and accessible to the operator. Long cables are
provided to connect to the Capnograph/Ventilator Interface Module.
Ventilator operation
Refer to the MRI-1 Ventilator Instruction Manual for detailed instructions for connecting
and using the ventilator.
microCapStar Capnograph
A miniature, infrared CO2 sensor with a temperature controlled sample cell prevents
water condensation even during long measurement sessions. Low sample flow and
rapid response is achieved with a carrier gas system where the high flow carrier gas
quickly brings the low flow sample to the sample cell.
Capnograph setup in MR

Table of Contents