Appendix C
Troubleshooting
Revision C, August 2013
C-1
No waveform on
display. Trace
frozen &
communication
error.
Wrong com port
selected.
Select the correct com port in
PC-SAM setup menu.
2-3
3-6
No waveform on
display. Trace
frozen & no
communication
error.
No power to the ERT
Control/Gating Module
or communication
problem.
Check the power and serial
connections
2-1
2-2
but no ECG signal.
the ERT Module.
Module & fiber optic cable
connections. Look for red light
in the fiber optic driver.
2-2
but no ECG signal.
Lead off.
Check lead attachment.
4-2
but no ECG signal.
Offset voltage too large.
Check offset voltage in system
info window. Check for faulty
temp probe by disconnecting
the probe from the ERT
Module & check for needles in
muscle.
9-1
4-7
no heart rate
display.
incorrect.
on ECG without
scanning’
magnet bore.
Check for vibration or air flow.
Tape leads and/or ERT
module.
4-1
on ECG while
module in magnet bore.
Check for vibration. Tape leads
& ERT module to restrain
4-1