Appendix A
Specifications
Revision C, August 2013
A-1
ERT:
ECG Range: 40 - 900 BPM
Accuracy: ±1%
Input range: -2.50 mV to 2.5mV
Input Impedance: >10 M @ 10 Hz
CMRR: 100 dB @ 60 Hz
Resp Range 15 - 300 bpm
Accuracy 1 count
Sensor pneumatic pillow
and/or ECG lead
Temp Probe type thermister, rectal
Tip OD, mm 1.0 and 2.5
Range 0 100 ºC
Accuracy +/-0.2 ºC, 1070ºC
Module Power - battery Rechargeable
Battery life: >15 hours
Time to full charge <2 hours
ERT module size: hxwxd cm 2.1x5.1x15.0
Battery Pack size: hxwxd cm 1.8x5.6x7.4
ERT Control/Gating:
Gating R-wave to gate delay selectable - 0 to 600 ms
Expiration gate width selectable - 1 ms
and delay step size
Temp Heater control fiber optic PWM
Size: hxwxd cm 3.8x13.3x12.5
Module Auxiliary inputs 2 TTL
Power 12 VDC
Air Heater System:
Heater Heater control fiber optic PWM
power 100/115/ 230 VAC
Size: hxwxd cm 16.4x8.9x26.0
Fan power 100/115 /230 VAC
Size: hxwxd cm 6.4x18.0x18.4
Water Bed Heater System:
Heater Heater control mixed water temperature
35 45 ºC ±0.1 ºC
power 100/115/ 230 VAC
Size: hxwxd cm 16.4x8.9x26.0
Circulation power 12 VDC, 2.0 A
Size: hxwxd cm 32x54x33