Closing the Combics indicator:
t Re-attach the front panel and secure it with 1 Nm the ten cap nuts.
Connecting the Device to AC Power
The device is powered through the installed power cord. The power supply is
integrated into the indicator. The device can be operated with a voltage of 100 V
to 240 V.
3
The power connection must be made in accordance with the regulations applicable
in your country.
The printed voltage rating (see type label) must match the voltage in the place of
installation. If the voltage specified on the label or the plug design of the AC adapter
do not match the rating or standard you use, please contact your Sartorius office or
dealer.
t Check the voltage rating and plug design.
t The device must be plugged into a properly installed wall outlet.
Protection Class 1 Device
t The device must be plugged into a properly installed wall outlet which has a
protective grounding conductor (PE).
Safety Precautions
3
If you use an electrical outlet that does not have a protective grounding conductor,
ensure that an equivalent protective conductor is installed by a certified electrician
(as specified in the applicable regulations for installation in your country).
The protective effect must not be negated by using an extension cord without
a protective grounding conductor.
Before using for the first time, any superstructure parts must be completely installed.
Avoid connecting the equipment to lines that have a heavy electrical load, e. g.
compressors, large machinery, etc.
Warm-up Time
To deliver exact results, the device must warm up for at least 30 minutes after
connection to AC power. Only after this time will the device have reached the
required operating temperature.
Using a Verified Device in Legal Metrology:
Ensure that there is a warm-up time of at least 24 hours after connection to the
power supply.
Operating Instructions Combics Indicators 15
Getting Started

Table of Contents

Other manuals for Sartorius combics 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 2 and is the answer not in the manual?

Sartorius combics 2 Specifications

General IconGeneral
BrandSartorius
Modelcombics 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals