4
– The 609308-..1 AC adapters or
power supplies are designated
[EEx ib]IIC and II (2) G and may
only be installed as an additional
electrical apparatus outside the
hazardous area/location
(approval certificate no. KEMA
98ATEX2752X
All AC adapters or power supplies
may only be connected to an electrical
outlet (mains supply) (48 – 62 Hz) with
at least 184 V and no more than
264 V.
The explosion-protected equipment of
the Economy Series meets the currently
valid, harmonized European standards
in compliance with CENELEC:
EN 50014: 1992
EN 50020: 1994
The flameproof AC adapter or power
supply is also in compliance with:
EN 50018: 1994
The Economy Series scales and any
accessories or auxiliary equipment
must be installed and put into
operation according to the
acknowledged technical regulations.
You must also comply with the national
laws and safety regulations of your
country.
Important Note to Users !
Make sure to carefully read and follow
sections marked with this symbol –
they contain important safety
instructions.
Warning and
Safety Information
Improper use or handling can result
in property damage and/or personal
injury. To prevent damage to the
equipment, read these operating
instructions thoroughly before using
your scale. Keep these instructions
in a safe place.
The manufacturer shall not be held
liable for damage caused by failure
to observe this warning and safety
information or the instructions.
When you use electrical equipment
in installations and under ambient
conditions requiring higher safety
standards, you must comply with the
provisions specified in the applicable
regulations for installation in your
country. Before you operate your
scale in a hazardous area/location,
the scale must be inspected either by
a certified electrician or under the
guidance and supervision of a
certified electrician to make sure that
the scale complies with the applicable
regulations (in Germany, Section 12 of
the ElexV). Determine whether your
scale must be reported to the technical
inspection authorities (such as the trade
board) in your country. The system
must also be inspected during
operation. The system should be
inspected at intervals which allow for
early detection of the faults which
occur as a result of normal wear and
tear, so that they can be corrected
before damage is caused. In any
case, inspection must be performed
at least every 3 years.