69
Pin Assignment Chart
Female Interface Connector:
14-position round connector with screw lock
hardware for cable gland
! Warning When Using Pre-wired RS-232
Connecting Cables
RS-232 cables purchased from other manufacturers
often have incorrect pin assignments for use with
Sartorius scales. Be sure to check the pin
assignment against the chart below before
connecting the cable, and disconnect any lines
marked “Internally Connected” (e.g., pin 6). Failure
to do so may damage or even completely ruin your
scale and/or peripheral device.