68
Software Handshake
The software handshake is controlled via XON and
XOFF. When a device is switched on, XON must
be transmitted to enable any connected device to
communicate.
Activating Data Output
You can define the data output parameter so that
output is activated either when a print command is
received or automatically and synchronous with the
scale display or at defined intervals (see application
program descriptions and auto-print setting).
Data Output by Print Command
The print command can be transmitted by pressing
p or by a software command (Esc P).
Automatic Data Output
In the “auto print” operating mode, data is output
to the interface port without a print command.
You can choose to have data output automatically
at defined print intervals with or without the stability
parameter. Whichever parameter you select, the
data will be output as the readouts appear on
the scale display. The display update frequency
depends on the setting for “Adapting the Filter”
(1 1 x).
If you select the auto print setting, data will be
transmitted immediately the moment you turn on
the scale.