Getting Started
Cubis MSE Operating Instructions 15
Connecting the Weigh Cell to the Electronics Module/Evaluation Unit
t Connect the weigh cell and electronics module using the connection cable�
Make sure that the connectors click into place so that both ends are securely
connected�
3
Please check the plug contacts to ensure a proper connection�
There should be no tension on the connection cable�
We therefore recommend placing the device close to the wall�
t Connecting the Weigh Cell to the Evaluation Unit
Make sure that the connectors click into place so that the connection is secure�
3
Models with a separate electronics module (e�g� MSE6�6/3�6/2�7),
MSE225S/P and MSE125P: Do not replace the electronics module�
Replacing them with models from the ME/SE series, for example, can
damage the device�
Microbalances (Models MSE6.6S/3.6P/2.7S):
Placing Components on the Weigh Cell
t Install the following parts in the order listed:
1� MSE2,7S models: bush
2� Shield plate
3� Weighing pan
Note: After inserting the weighing pan, turn it slightly to the left and right,
while pressing it down lightly�
4� Internal draft shield (for MSE2�7S only)
5� Draft shield: Center the hole over the pan (see arrows)
Filter Balances (Models MSE6.6S-F/2.7S-F):
Placing Components on the Weigh Cell
t Place the components listed below inside the weigh cell in the order given:
1� MSE2,7S-F models: female connector
2� Shield plate
3� Internal draft shield
4� Filter weighing pan d 50 mm or weighing pan
(optional filter weighing pan d 75 mm or d 90 mm)
Note: After inserting the pan, turn it slightly to the left and right,
while pressing it down lightly�
5� Draft shield cover
!
Tip: Turn the balance off and then on again after replacing the pan during
operation�
3
2
4
5
1
MSE6�6S/3�6P/2�7S:
MSE225…, MSx125:
MSE6�6S-F/2�7S-F:
MSE6�6S/3�6P/2�7S: