Care and Maintenance
76 Cubis MSE Operating Instructions
Care and Maintenance
Service
Regular servicing by a Sartorius technician will extend the service life of your balance
and ensure its continued weighing accuracy� Sartorius offers its customers service
contracts with regular maintenance intervals ranging from 1 month to 2 years�
The frequency of the maintenance intervals depends on the operating conditions
and user‘s tolerance requirements�
Repairs
Repair work must only be carried out by trained service technicians� Repairs
performed by untrained persons may result in considerable hazards for the user�
The device should be unplugged during repair work�
3 Unplug the power cord from the outlet�
Repair work must only be performed by Sartorius-trained service technicians�
If required: Please contact your Sartorius dealer!
Cleaning the Balance
Ensure that no dust or liquid enters the balance weighing system�
3 Do not use aggressive cleaning agents (solvents or similar�
Isolate from supply voltage: Unplug the power cord from the wall outlet and
unplug any data cables connected to the balance�
Make sure all equipment operators have received proper instructions when working
with hazardous or toxic materials�
Failure to follow correct handling procedures can result in breakage of parts or
spillage of liquids or other substances�
Make sure the necessary protective clothing or equipment is worn, such as
protective gloves, clothing, eyewear�
All safety regulations applicable in the laboratory must be observed�
When using the balance in the chemical industry, handle defective or device parts
to be cleaned according to respective regulations�
Models with a Readability of >1 mg:
– These models meet the requirements of protection class IP54�
The IP protection only applies to the balance housing and not to the AC adapter�
– The IP protection is only guaranteed when the weighing pan is mounted�
– The IP protection is limited when the data interfaces are used� Retain the cover
cap for the data output� When the data output is not in use, reapply the cover
cap to protect it against vapors, moisture, and dust or dirt�
t Open the panels of the analytical draft shield completely�
t Clean the housing and interior of the balance with a cloth lightly moistened
with a soap solution�
t Dry all parts with a soft, dry cloth or use blotting paper to absorb dampness�
Then replace all dried parts�
t Remove panels from the analytical draft shield and clean them with
commercially available glass cleaner� Dry the panels�
Then reinstall them to the balance�