Page #47 background imageLoading...
Page #47 background image
Cubis MSE Operating Instructions 47
Application Programs
Parameter configuration: Applic. - Density - dec.plcs. -1 dec.pl.
Example: Determining the density of a solid sample� The density at 20�0 degrees Celsius is 0�99823 g/cm
3
Step Press key Display/Printout
1� Attach sample holder and suspension wire
2� Tare the balance J 0.0 g
3� Change density of buoyancy liquid S _1.00000
4� Set density value (in this example: 0�99823)
Enter numerals in increments of 1 or: Repeatedly press S, press
and hold the V key _0.99823
5� Set the density value and start the application� V
The current density value will be saved in protected
memory until the setting is changed�
6� Confirm “air” display V air
7� Determining the weight of the sample in air:
Place sample on the balance + 20.0 g
8� Save weight in air V
9� Remove sample from the balance water
10� Determining weight value in liquid:
Place sample in basket
11� Confirm “water” display V 0.0 g
12� Place sample in liquid + 15.0 g
13� Save weight value in liquid, display and print results V
+ 4.0
o
RhoFl 0.99823 o
Wa + 20.0 g
Wfl + 15.0 g
14� Delete results F Rho 4.0 o
15� Repeat as needed, starting from Step 5

Questions and Answers: