Application Programs
Cubis MSE Operating Instructions 51
Example: Determining residual weight in percent
Parameter settings: Applic. - percent
Reference percentage value: Ref 100%
Step Press key Display/Printout
1� Tare the balance J 0.0 g
2� Information:
Enter the reference percentage S ref 100
%
(Changing the reference: see the previous page)
3� Place sample equal to 100% on the balance
(in this example: 111�6 g)
4� Initialize the balance V + 100.0
%
The current reference weight remains pRef 100 %
saved until a new reference is set Wxx% + 111.6 g
5� Remove sample (e�g� for drying)
6� Place unknown weight on balance + 94.9
%
(in this example: 105�9 g)
7� If desired, print the percentage P Prc + 94.9 %
8� Toggle display between weight and percentage + 105.9 g
Repeatedly press S + 94.9
%
9� Clear display of residual weight and reference percentage F + 105.9 g
10� If desired, print the residual weight P N + 105.9 g

Questions and Answers: