Application Programs
Cubis MSE Operating Instructions 53
Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example, with surface dimensions of 0�210 m
+ 0�297 m = 0�06237 m
2
� To determine the weight per unit area, the total weight is divided by the surface�
Parameter settings:
Applic. - calc. - method - Divis.
Step Press key Display/Printout
1� Tare the balance J 0.00 g
2� Activate divisor input S ______1.0
3� Set the divisor (in this example: 0�06237):
Position the decimal point, V, 4+ S, ___.00000
2+ V,
Enter numerals in increments of 1 or: Repeatedly press S ___.06000
briefly or hold
the, V key ___.06237
4� Save the divisor and initialize the balance V + 0.0
o
The current divisor will be saved in protected Div 0.06237
memory until the setting is changed�
5� Determine the weight per unit are: + 79.7
o
Place a sheet of DIN A4 paper on the balance
6� If desired, print the result P Res + 79.7 o
7� Toggle display between weight and calculated value Repeatedly press S + 4.97 g
+ 79.7
o
8� Unload the balance + 0.0
o
9� Repeat as needed, starting from Step 5