Application Programs
Cubis MSE Operating Instructions 55
Example: Automatic weighing of animals with 20 subweighing operations
Parameter configuration: Applic. - AnimalW.
Step Press key Display/Printout
1� Place animal weighing bowl on the balance 22.6 g
2� Tare the balance J 0.0 g
3� Change the number of subweighing operations S ref 30
4� Set number of measurements: Repeatedly press S
In increments of one (1, 2, 3, ���, 100) briefly or
In increments of ten (10, 20, ���, 100) hold the S key ref 20
5� Confirm selected subweighing operations V + 0.0 g
and start automatic animal weighing�
The number of subweighing operations will be saved
in protected memory until the setting is changed�
6� Place first animal in bowl� 888
The balance delays the start of measurements 20
until the difference between 4 measurements 19
meets the criterion� ....
1
7� Read the result + 410.1 g G
The result is displayed with the “G” (= calculated value)
and remains displayed until the sample (animal) mDef 20
is removed from the load plate (bowl)� x-Net + 410.1 g
8� Unload and tare the balance + 0.0 g
9� Weigh next animal (if desired)�
The next weighing series starts automatically�