Cubis MSE Operating Instructions 89
Technical Specifications
Model-specific Data
Semi-microbalances 0�01 mg
Model MSE225S MSE225P MSE125P
Readability mg 0�01 0�01/0�02/0�05 0�01/0�1
Weighing capacity g 220 60/120/220 60/120
Tare range (subtractive) g – 220 – 220 – 120
Repeatability <±mg 0 to 60 g: 0�015 0 to 60 g: 0�015 0 to 60 g: 0�015
60 to 220 g: 0�025 60 to 220 g: 0�04 60 to 120 g: 0�06
Linearity <±mg 0�1 0�15 0�15
Corner load (test load [g]) mg 0�15 (100) 0�2 (100) 0�15 (50)
Optimal starting point
of the measuring range* mg 8�2 8�2 8�2
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s < 2 < 2 < 2
Typical measurement time s < 6 < 6 < 6
External standard calibration value g 200 (E2) 200 (E2) 100 (E2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�2 – 0�4
Weighing pan size (W + D) mm 85 + 85
Weighing chamber height (draft shield DU)
mm 261
Protection Protected against dust and water
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
Analytical balances 0�1 mg
Model MSE524S MSE524P MSE324S MSE224S MSE324P MSE124S
Readability mg 0�1 0�1/0�2/0�5 0�1 0�1 0�1/0�2/0�5 0�1
Weighing capacity g 520 120/240/520 320 220 80/160/320 120
Tare range (subtractive) g – 520 – 520 – 320 – 220 – 320 – 120
Repeatability <±mg 0�1 0�15/0�2/0�4 0�1 0�07 0�1/0�2/0�4 0�1
Linearity <±mg 0�4 0�5 0�3 0�2 0�5 0�2
Corner load (test load [g]) mg 0�3 (200) 0�4 (200) 0�3 (200) 0�2 (100) 0�4 (200) 0�2 (50)
Optimal starting point
of the measuring range* mg 82 82 82 82 82 82
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1 1 1 1
Typical stabilization time s < 1 < 1 < 1 < 1 < 1 < 1
Typical measurement time s < 3 < 3 < 3 < 3 < 3 < 3
External standard calibration value g 500 500 200+100 200 (E2) 200+100 100 (E2)
(min� accuracy class) (E2) (E2) (E2) (E2)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 85 + 85
Weighing chamber height (draft shield DU)
mm 261
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�