8 Specifications
8.1 General Data
Specification Unit Value
Scale, ID code:
(explosion protection) II 2G Ex ib IIB T4
Gb as per EC Type Examination
Certificate no. FM14ATEX0085X
Power supply only via USB interface or Sartorius AC adapter YPS06-USB
Input supply voltage V
DC
+5.0
Power consumption W 2.0
Other data IP40 in accordance with EN 60529/IEC 60529
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment For indoor use only
Operational capability °C Guaranteed between +5 and +40
Storage and shipping °C -10 to +60
Relative humidity % 15% to 80% for temperatures up to 30°C non-condensing,
decreasing linearly to 50% relative humidity at 40°C
Ex-Link Converter Interface Connection USB, type B
Electromagnetic Compatibility As per EN 61326-1/IEC 61326-1 Electrical equipment for measurement,
control and laboratory use – EMC requirements – Part 1: General requi-
rements
Interference resistance Basic requirements
Transient emissions Class B
Suitable for use in residential areas and areas that are connected to a
low voltage network that also supplies residential buildings.
Available Application Programs Recalculation, factor calculation, formula
AC Adapter YPS06-USB
USB power plug (5 V/900 mA) Type FW7712 (manufacturer’s certificate)
Primary 100–240 V~, -10%/+10%, 50–60 Hz, 0.125 A
Secondary 5 V
DC
, ±5%, 900 mA (max.)
Other data Protection class II
IP40 in accordance with EN 60529/IEC 60529
Ex-Link Converter YCO14-Z
ID code (explosion protection)
II (2)G [Ex ib Gb]IIB as per EC Type Examination Certificate no.
FM14ATEX0084X
Other data IP40 in accordance with EN 60529/IEC 60529
10 EVO1X | LAB1X Installation Instructions