EasyManua.ls Logo

Sartorius Portable PT 210-000V1 - Adjusting the Balance;Scale; Adjustment Sequence; Opening the Balance;Scale Housing

Sartorius Portable PT 210-000V1
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
Adjusting the Balance/Scale
Before you begin the adjustment procedure, let the balance/scale warm up for at least 30 minutes, and run a function
test.
Important Note:
The number(s) written next to the components, e.g., main PCB (118), refer to the exploded-view diagram on page 4.
Adjustment Sequence
Adjust the overload stops
Adjust the zero point offset value
Check the off-center load
Adjust the linearity
Calibrate the balance/scale
Opening the Balance/Scale Housing
Only if you wish to adjust the zero point offset value or the linearity will you need to open the balance/scale housing.
All tests and other tasks can be performed with the balance/scale housing left closed.
- Disconnect the balance/scale from the operating voltage and remove the
weighing pan (101).
- If the balance/scale has a rectangular weighing pan, you must first remove
the screw (103.A) along with the leaf spring for grounding (102), and the
pan support (103).
- Turn the balance/scale over and place it with the pan side facing
downward on a padded surface to prevent damage to the weighing
system.
- Remove the 3 screws that are now accessible (115.A) from the housing
base(115).
- Place the balance in the upright position; carefully remove the upper
housing section(104) and place it at the left beside the balance/scale.
- Now you can access all components located on the main PCB (118)
required for adjusting the zero point offset value and the linearity.
- Follow these steps in reverse order to close the balance/scale housing.

Related product manuals