65
R
8. Compare the result to the range printed on the control solution insert. If the results are not
within the lipid control range printed on the insert, then the Analyzer and test strip may not
be working properly. Repeat the control test.
If the control solution result is still not acceptable, please contact SD Biosensor, Inc. service
center at +82-31-300-0400.
9. Remove the used the test strip for control solution from the Analyzer and discard it.
CAUTION
-
This solution is made by human and animal source materials. Treat as potentially infections.
- The human source material used to produce this product has been tested by using FDA-
accepted methods and found nonreactive for hepatitis B surface antigen (HBsAg), and
for antibodies to hepatitis C (HCV) and human immunodeficiency viruses(HIV-1 and
HIV-2). Because no test can offer complete assurance that infectious agents are not
present, this product should be considered potentially infectious and handled with the
same precautions used with patient specimens.
- Store upright and refrigerated at 2-8˚C (36-46˚F). Stored under this condition, SD Lipid
Control Solution can be expected to give stable results through the expiration date prineted
on the
label. Minimize exposure to strong light.
- Do not use SD Lipid Control Solution followings :
t If they are cloudy or have an odor
t If they have been shipped or stored improperly