CHAPTER 10.
Warnings, Precautions and Limitations
Discard test strip after using. Strips are to be read once. Never insert or read a used test strip.
Do not ingest.
Discard the used test strip according to the local guideline.
1. Test Strip for Lipid
1. Make sure the codechip number is matched with the lipid test strip code number. Never use
a codechip from a different lot.
2. Out-of-date or expired strips cannot be used in your test system. Check the expiration date
in the package or pouch.
3. Add all of the blood to the lipid test strip at one time. If you do not get all of the blood on
the strip, do not add blood to the same strip. Test again with a new test strip and fresh blood
sample.
4. Venous whole blood, serum and plasma is only for professional use.
2. Test Strip for Glucose
1. Use only fresh capillary blood. Do not use serum or plasma or venous whole blood.
2. The SD LipidoCare® System is not designed to be a substitute for pathology laboratory
equipment and should not be used for the diagnosis of diabetes.
3. Extremes in hematocrit may affect test results. Hematocrit levels less than 20% may cause
falsely high readings. Hematocrit levels greater than 60% may cause falsely low readings.
4. Not intended for screening or diagnosis or for use on neonates.
5. Never make significant changes to your diabetes control program or ignore physical
symptoms without consulting with your healthcare professional.