67
R
CAUTION
t The SD Glucose Control Solution range printed on the glucose test strip container is
for SD Glucose Control Solution only. It is not a recommended range for your blood
glucose level.
t Store SD Glucose Control Solution tightly closed at temperatures between 8-30°C
(46-86°F ). Do Not refrigerate.
t SD Glucose Control Solution, should be tested at room temperature 18-30°C(64-86°F).
t If a glucose control solution test is still not acceptable, please contact SD Biosensor,
Inc. service center at +82-31-300-0400.
6. Compare glucose control solution result to the range printed on the glucose test strip
container. If the results are not within the glucose control range printed on the glucose test
strip container, then the Analyzer and glucose strips may not be working properly. Repeat
the SD Glucose control solution test. If the control solution result is still not acceptable,
please contact SD Biosensor, Inc. at +82-31-300-0400.
7. Remove the used glucose test strip for control solution from the Analyzer and discard it.