77
R
Indication Error description
In case of lipid test strip
In case of glucose test strip
Error: blood sample Error
An insufficient amount of blood was applied.
Solution
Discard this test strip and test again using a new test strip with
a larger sample, making sure blood is placed to the narrow
channel in the top edge of the test strip.
Indication
Error
description
Error: Expired strip (Only for a Lipid test strip)
The Lipid test strips are expired.
Solution
Discard this test strip and test again using a new lipid test strip
does not past expiration date.
Indication
Error
description
Error: Temperature Error
If the environmental temperature is above or below the
operation range of an Analyzer, a thermo Analyzer icon will
appear on the display.
Solution
Move to an area at 18-32°C(64-90°F) for lipid or 10-45°C (50-
113°F) for glucose, wait for 30 minutes, and perform a test. Do
not artificially heat or cool the Analyzer.