directly(if not, please sweep over SET/SELECT until it points to FCF); press SET until
indicator light turns red & blinks; input new FCF( refer to Step 4); done;
6. Press SELECT until comes pop-up menu, asking whether to return to main interface;
sweep over SET for yes, SELECT for no, and sweep over SET/SELECT again to confirm.
Every meter has its own calibration coefficient marked on standardization certificate and
nameplate of sensor. Each sensor has its own paired transmitter. Customer can use the
meter directly without setting new FCF, but in the following 2 cases: 1. Transmitter is
replaced; 2. Measuring result has a big difference from real value.
Formula:
● By software
Start the software; check tool bar on the top; click Functions --> Configuration; comes setting
interface; click Modified on the right of FlowCalFactor; input new FCF; click OK to confirm;
click Application; done.
7.2 Min. flow cut-off setting
1.On main interface, sweep over screen as Pic. 7-1 until comes interface ;
2.Sweep over SET/SELECT as Pic. 7-2 until cursor points to ;
3.Press SET until indicator light turns red & blinks, comes the password input interface;
4.Input password:
a. Sweep over SET/SELECT to set the value, SET=Decrease, SELECT=Increase.
b. Sweep over screen as to move to nearby digit, from SET to SELECT=Move right,
SELECT to SET=Move left.
c. Press SET until indicator light turns red & blinks, to confirm password( or press SELECT
until indicator light turns red & blinks, to cancel operation).
5.Comes interface ; sweep over SET/SELECT until cursor points to
MFC(Min. flow cut-off); press SET until indicator light turns red & blinks; input new
MFC( refer to Step 4); done;
6.Press SELECT until comes pop-up menu, asking whether to return to main interface;
sweep over SET for yes, SELECT for no, and sweep over SET/SELECT again to confirm.