Zhejiang Sealand Technology Co.,Ltd.
● By software
Start the software; check tool bar on the top; click Functions --> Configuration; comes setting
interface; click Modified on the right of MassFlowCutoff; input new MFC; click OK to confirm;
click Application; done.
7.3 Pulse output frequency setting
1. On main interface, sweep over screen as Pic. 7-1 until comes interface ;
2. Sweep over SET/SELECT as Pic. 7-2 until cursor points to ;
3. Press SET until indicator light turns red & blinks, comes the password input interface;
4. Input password:
a. Sweep over SET/SELECT to set the value, SET=Decrease, SELECT=Increase.
b. Sweep over screen as to move to nearby digit, from SET to SELECT=Move right,
SELECT to SET=Move left.
c. Press SET until indicator light turns red & blinks, to confirm password( or press
SELECT until indicator light turns red & blinks, to cancel operation).
5.Comes interface ; sweep over SET/SELECT until cursor points to
MF(Pulse output frequency); press SET until indicator light turns red & blinks; input new
MF( refer to Step 4); done;
6.Press SELECT until comes pop-up menu, asking whether to return to main interface; sweep
over SET for yes, SELECT for no, and sweep over SET/SELECT again to confirm.
● By software
Start the software; check tool bar on the top; click Functions --> Configuration; comes setting
interface; click Modified on the right of MaxFrePulse; input new MF; click OK to confirm; click
Application; done.