Zhejiang Sealand Technology Co.,Ltd.
● By software
Start the software; check tool bar on the top; click Functions --> Fix Freq Out Test; comes
setting interface; click Fix Freq Output, the screen shows red spot (click UnFix Freq to cancel
this function, the screen shows green spot); click Modified to input fix frequency output.
7.7 Total clearance
There will be no record of current Total once cleared; thus, please keep a record manually if
necessary.
● By transmitter
There are 2 methods to clear Total.
Method 1
1.Sweep over screen as Pic. 7-1 until comes interface or ;
press SELECT until comes pop-up menu, asking whether to clear; sweep over SET for yes,
SELECT for no, as Pic. 7-2; sweep over SET/SELECT again to confirm; comes password
input interface;
2. Input password:
a. Sweep over SET/SELECT to set the value, SET=Decrease, SELECT=Increase.
b. Sweep over screen to move to nearby digit, from SET to SELECT=Move right,
SELECT to SET=Move left.
c. Press SET until indicator light turns red & blinks, to confirm password( or press
SELECT until indicator light turns red & blinks, to cancel operation); done.
Method 2
1. On main interface, sweep over screen as Pic. 7-1 until comes interface ;
2. Sweep over SET/SELECT as Pic. 7-2 until cursor points to ;
3. Press SET until indicator light turns red & blinks, comes the password input interface;
4. Input password:
a. Sweep over SET/SELECT to set the value, SET=Decrease, SELECT=Increase.
b. Sweep over screen to move to nearby digit, from SET to SELECT=Move right,
SELECT to SET=Move left.
c. Press SET until indicator light turns red & blinks, to confirm password( or press
SELECT until indicator light turns red & blinks, to cancel operation).