K600000000SE 030804 011018
CONTEXT KEY FUNCTION
Updates the measurement.
Saves, in the memory selected in the "Select Memory" menu, the data
acquired.
Starts printing the paper print-out. SEE SECTION 12.
13.4 Measurements→Ambient CO
Make sure to perform the autozero in a clean environment (preferably outdoors), so that
the ambient CO measurement is correct. It is advisable to turn on the instrument and wait
for the autozero completion outside the area where the test is being performed.
- The values of the ambient CO that you want to save must be acquired before saving the analyses.
- To join the values of the ambient CO to the measurements of the current analysis use the ' ' function.
- To print the paper print-out with the measurement of the ambient CO, activate the ' ' function
- It is possible to delete a draft value acquired by the memory by overwriting it by activating the ' ' function again.
- After saving the draft values, to carry out the combustion analysis, press the key ' '
15/01/14
10:00
Measurements
Ambient CO
CO
ppm
0
CO max
ppm
0
KEY FUNCTION
Activate the context keys shown on the display.
Returns to the previous screen.