73
K600000000SE 030804 011018
Warning: in automatic mode, the measurements of smoke, draft and ambient CO must be
taken before starting the combustion analysis.
Manual analysis mode
If the user chooses the manual mode, he will perform the combustion analysis manually;
in this case, the settings regarding printing and duration of the automatic analysis will not
be considered. At this point the user can start the manual analysis after waiting for the
measured displayed to stabilize: then the user can proceed to save or directly print the
data, which will be prepared in accordance with the previously configured settings.
At the end of the three analyses, the screen with the average can be displayed, which also
contains all the data necessary to fill in the booklet of the system or plant.
Memory selection mode
Manual: the memory will have to be selected manually via the parameter "Select"
Auto: the memory, to which the measurements and combustion analyses will be saved,
will be suggested automatically when the device is turned on.
SEE SECTION 11.5.
Allows the user to delete the contents of each memory or of the entire 99 memories.
SEE SECTION 11.6.
The user, through this menu, can view the percentage of memory usage.
SEE SECTION 11.7.
Data logger
Delete
Usage %