Section 2 SMT-200 Operations Io
SENSOR
30
SMT-200 Operations and Technical Manual 2.00/R1
2.24.4 File menu
Internal file handling and data / code transfer options
Redisplay Last
Displays the results of the last test run.
Storage status
Internal memory status and post mortem shutdown
status.
Dump data file
Send the data file to an external serial device (e.g. serial
printer or PC running in a terminal emulation mode).
List data file
List data file held in internal memory.
Erase data
Delete all records held in memory.
Read Config
Reloads SMT-200 configuration file from memory.
Save Config
Saves SMT-200 configuration file to memory.
2.25.4.1 Redisplay Last
Shows the result of the last test run
If a test has not been run since the SMT-200 was switched on there will be
no meaningful data shown.
File menu
----------------
1>Redisplay last
2 Storage status
3 Dump data file
4 List data file
5 Erase ALL data
6 Read Config
7 Save Config
0 Back to Main
----------------
Redisplay the
Last test data
4>File menu...
4>File menu...
1>Redisplay last
ay Last
----------------
No. 14
Temp 21.6
o
C
Pol 1
Leak 9.2 Mohm
Res 394 ohm
Freq 10.35 Hz
Damp 0.242
Sens 28.83 V/m/s
Dist 0.10 %
Imp 2393 ohm
Test OK
----------------