10
1. Low Power Cartridge or warm Power Cartridge
Eclipse is: Beeping, has a YELLOW LED flashing and the unit IS running.
Å Continued from Previous Page
C. Is Eclipse running from the AC or DC Adapter?
i. AC Adapter
There may have been an electrical fluctuation (an electrical dip or
spike), or a power outage at the electrical outlet, Eclipse has switched to the Power
Cartridge as a backup, and now the Power Cartridge is either low or warm (see
Steps B i or B ii)
ii. DC Adapter
The voltage in the vehicle has possibly fallen out of range and
Eclipse has switched to the Power Cartridge as a backup, and now the Power
Cartridge is either low or warm (see Steps B i or B ii)
D. Press the No Smoking Icon once, and observe what number appears after the “AC=” in
the LCD window.
(The Diagnosis Code will disappear after 5 sec’s if no button is pushed).
** NOTE: If Eclipse is plugged into an external power source and running from the Power Cartridge
and will not switch over to an external power source (AC or DC) then there may be a system
fault. Exchange out the unit for evaluation.