Quick Reference Alarm Guide
Alarm Priority Green Yellow Red Alarm Code
All OK, warm-up done 13/None
ON OFF OFF OFF 00
Warm-up: %O
2
< 85 12/Low ON ON OFF OFF 08
Warm-up: %O
2
< 70 11/Low ON ON ON OFF 04
Cannot Charge Battery 10/Low
N/A ON OFF OFF 01
No breath for 45 seconds 9/Medium
ON ON OFF OFF 20
70% < %O
2
< 85% 8/Medium OFF (1s/1s) OFF OFF 08
Flow rate error 7/Medium OFF
(1s/1s) OFF (0.5s/119.5s) 20
Warm battery 6/Medium
ON (1s/1s) OFF (0.5s/119.5s) 02
Low battery 6/Medium
ON (1s/1s) OFF 2 Beeps 0.5s/0.5s every 30s 10
DC cable detected 4/Medium
N/A (1s/1s) OFF (0.5s/1.5s) N/A
%O
2
< 70 3/High OFF OFF (0.25s/0.25s) 3 Beeps (0.5s/0.5 every 120s) 04
Loss of power 2/High OFF OFF
(0.25s/0.25s) ON 40
Malfunction 1/High OFF OFF
ON 10s then silence N/A
Troubleshooting Guide
Troubleshooting Guide