11436 Sorrento Valley Road • San Diego, CA 92121 • www.sequal.com • 858.202.3100
20
Troubleshooting Guide
Troubleshooting Guide
9. Power Cartridge will not charge more than 5 bars (E1) & more than 2 bars in
E2.
Eclipse shuts down at half the time expected while running on the Power Cartridge.
A. Has the Power Cartridge and display been calibrated recently? If “YES”, then go to Step vi
i. If “NO”, Make sure the Power Cartridge is plugged into the Eclipse.
ii. Remove external power source (AC or DC Adapter) if present.
iii. Continue to run Eclipse on the Power Cartridge until the Eclipse STOPS RUNNING,
with a LONG CONTINUOUS ALARM and the Red LED “Flashing”
NOTE: Approximately approx 10 min’s before the Eclipse Power Cartridge is completely discharged, the
Low Battery alarm will sound (1 beep every 30 sec’s) and a Yellow Flashing LED will come on.
iv. Plug in the AC Adapter into the Eclipse’s power connector, once the Battery is
completely discharged.
v. Allow the Eclipse to charge the Power Cartridge until the Power Cartridge Indicator is
completely.
* NOTE: By powering down or reducing the liter flow the Power Cartridge charge time will be reduced.
vi. If “Yes”: then the Power Cartridge maybe faulty. “Exchange the item for evaluation and
possible repair.”