9
1. Low Power Cartridge or warm Power Cartridge
Eclipse is: Beeping, has a YELLOW LED flashing and the unit IS running.
A. Is Eclipse plugged into an External Power Source or running off of the Power
Cartridge?
i. If POWER CARTRIDGE, then go to Step B
ii. If External POWER SOURCE (AC or DC), then go to Step C
B. How many bars are on the Power Cartridge Icon?
i. 1 bars or Less
i. The Power Cartridge power is low in the Eclipse. Replace the Power Cartridge
with a fully charged one, or switch to an external power source (Use AC or DC
Adapter).
ii. 2 Bars or More
i. The Power Cartridge is possibly too warm. Replace the Power Cartridge with a
spare one (that has cooled), or switch to an external power source (Use AC or
DC Adapter), and remove the Power Cartridge for approximately 15 to 20
minutes to allow to cool properly, before replacing.
** NOTE: If Eclipse is plugged into an external power source and running from the Power Cartridge
and will not switch over to an external power source (AC or DC) then there may be a system
fault. Exchange out the unit for evaluation.