LISST-200X User’s Manual LISST-200X Operation Details Page 73
The time delay start condition will cause the program to wait the
specified number of seconds before continuing on to the data collection
routine.
The LISST-SOP software can configure the sampling program with one
of six Stop conditions: Depth, Time. External Mechanical Switch,
External Digital Input, Fixed number of samples, and Maximum
Memory or Low Battery. The details of each condition are described
below.
The LISST-SOP software can configure the LISST-200X with one of six
stop conditions:
1. Depth,
2. Time,
3. External Mechanical Switch,
4. External Digital Input,
5. Fixed number of samples, and
6. Memory full.
When storing data in the Fixed Sample Rate mode the Stop conditions
are checked after each averaged sample has been saved. When
storing data in the Burst mode the Stop conditions are only checked
after a full Burst has been completed. The Start and Stop conditions
have no effect on the Real-time sampling mode. The details of each
condition are described below.
The depth sensor of the LISST-200X is used to check the current depth
to determine if it is less than the desired Stop depth. If the depth is less
than the threshold the sampling will stop. If the Start Condition is Depth
Start the program will return to checking for the start depth. For all
other Start Conditions the when the current depth is less than the
threshold the LISST-200X will stop and go into deep sleep.
The program will check the current time after each sample or burst and
compare it to the Start Time. If the Start Time is equal to or later than
the current time, it will stop and go into deep sleep.