LISST-200X User’s Manual LISST-200X Operation Details Page 74
External
Mechanical
Switch Stop
After each sample or burst the status of the Switch lever is checked. If
the switch lever is in the off or “0” position sampling will stop. If the
Start Condition is a Switch Start the program will return to checking the
start condition. For all other Start conditions the instrument will stop and
go into deep sleep.
External Digital
Input Stop
The status of the digital 1 input is checked after each sample or burst. If
the voltage at the input is less than 0.7 volts the sampling will stop. If
the Start Condition is a Digital Input Start the program will return to
checking the start condition. For all other Start conditions the
instrument will stop and go into deep sleep.
Fixed Number of
Samples
The program will acquire a fixed number of samples before stopping.
When the number of samples to be saved has been reached the
program will stop and go into deep sleep.
Memory Full
Sampling will continue until the memory is full, or the power input falls.
If the memory fills, the instrument will go into deep sleep.