Servomex 2700 Quickstart Manual
Servomex 2700 Quickstart Manual
30
Hazardous area classification
Sensor head Non hazardous locations.
Control unit Approved as Ex ic nA nC IIC T5 Gc, Ex tc
IIIB T75°C Dc (Ta = -10°C to +55°C) for
ATEX Group II, Category 3,
Gas and Dust hazardous atmospheres,
certificate number ITS10ATEX47005X.
Approved as non-incendive for:
Class I, Div. 2, Groups A,B,C & D.
Class II, Div. 2, Groups F & G.
Class III, Div. 1 & 2.
Enclosure Type 4X.
T5. Ambient Temperature 55°C max.
(Factory Mutual - FMRC)
Suitable for use in:
Class I, Div. 2, Groups A,B,C & D.
Class II, Div. 1, Groups E,F & G.
Class III, Div. 1.
T5. Ambient Temperature 55°C max.
Enclosure Type 4X.
(Canadian Standards Association - CSA)
Design standards
Conforms to the following standards:
Electromagnetic Compatibility
IEC 61326-1: 1997 (BS EN 61326-1: 1997)
Control unit - Emission Class B
Sensor head - Emission Class A
Electrical Safety
EN61010-1: 1993
IEC1010-1: 1990: amendments A1:1992 and A2:1995
WARNING
To minimise the risk of propagating brushed discharges, do
not install in a high velocity dust laden environment.