EasyManuals Logo

Servomex DF-150E User Manual

Servomex DF-150E
88 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
14 DF-150E Analyzer Start-Up Procedure
regulator to set the flow rate.
NOTE
For best results, after putting electrolyte in the sensor, allow the
sensor to sit for 60 minutes before starting gas flow.
5.2.1 Gas Connection
1. Connect the gas line to the
1
/8 inch compression fitting marked sample inlet
on the rear of the analyzer.
2. Open the door and slowly adjust the flow rate to 1.0 scfh as indicated on
the flowmeter using the flow control valve or pressure regulator. Do not
control the flow with the valve mounted in the flowmeter on positive
pressure installations.
3. If you are using a gas other than nitrogen, you must set the “indicated”
flow rate as shown in Table 1.
Background Gas Flowmeter Setting (scfh)
Argon 1.2
Ethylene 1.0
Nitrogen or Air 1.0
Carbon Monoxide 1.0
Methane 0.8
Hydrogen 0.3
Helium 0.4
Table 1: Flowmeter Settings versus Background Gas
NOTE
Unless otherwise requested, the analyzer is factory calibrated on
nitrogen. Analyzer readings will be inaccurate if used with a
background gas other than the one for which it is calibrated. If the
analyzer was calibrated on a gas other than nitrogen, the gas will
be noted on the calibration log inside the front door and on a label
below the display.
5.3 Making Output Signal Connections
The following considerations are important when performing this step:
The analog output signal is factory-set to 0 to 10 VDC unless otherwise
specified.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Servomex DF-150E and is the answer not in the manual?

Servomex DF-150E Specifications

General IconGeneral
BrandServomex
ModelDF-150E
CategoryAnalytical Instruments
LanguageEnglish

Related product manuals