49 | Page
MAINTENANCE
Continued next page
Temperature Calibration Continued
3
Place the display in temperature calibration mode.
a. Press and hold both the Up and Down temperature
arrow buttons simultaneously for approximately 5
seconds.
b. Release the buttons when the temperature display
shows the letters “C O”. The display will begin
flashing the current temperature display value.
Note: The display will automatically exit calibration mode after
5 seconds of inactivity, with the last shown temperature
display value saved.
4
Use the Up or Down arrows to adjust the current
display temperature value until it matches the
reference device temperature reading.
Reference Device
5
After matching the display to the reference device, wait 5
seconds.
The temperature display will cease flashing
and store the corrected chamber display value.
The incubator will now begin heating or
passively cooling in order to reach the setpoint
with the corrected display value.
Cooling to Setpoint
6
Allow the incubator to operate for at least 1 hour
undisturbed to stabilize after the incubator has
achieved the corrected temperature setpoint.
Failure to wait until the incubator is fully
stabilized will result in an inaccurate reading.
Setpoint Achieved