MAINTENANCE
End of Procedure
Temperature Calibration Continued
7
Compare the reference device reading with the chamber
temperature display.
If the reference device and the chamber temperature
display readings are the same or the difference falls
within the range of your protocol, the incubator is now
calibrated for temperature.
-OR-
See the next step if the readings fail to match or fall
outside of your protocol range.
Reference Device
8
If the difference still falls outside the acceptable range of your
protocol, repeat steps 3 – 7 up to two more times.
Reference Device
9
If the temperature readings of the incubator temperature display and the reference device
still fall outside your protocol after 3 calibration attempts, contact your distributor or
customer support for assistance.