EasyManua.ls Logo

Sheldon Shel Lab - Page 52

Sheldon Shel Lab
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52 | Page
MAINTENANCE
5. The CO
2
concentration in the incubation chamber must be stable in order to perform an accurate
calibration.
The unit must be thermally stable, having operated at your application temperature for at
least 12 hours prior to starting the gas calibration.
After turning on a flow of CO
2
to the chamber and setting the CO
2
setpoint, allow the
incubator to operate undisturbed with the chamber door shut for at least 2 hours.
To be considered stabilized, the incubator chamber must operate at your calibration CO
2
concentration for at least 30 minutes with no fluctuations of ±0.1% or greater.
Continued next page
A Suggested CO
2
Calibration
1
Once the incubator CO
2
concentration has stabilized, compare the
reference device and the CO
2
display readings.
If the readings are the same, or the difference
between the two falls within the acceptable range of
your protocol, the display is accurately showing the
incubation chamber CO
2
concentration. The CO
2
Calibration procedure is now complete.
-OR-
If the difference falls outside of your protocol range,
advance to Step 2.
Reference Device
Set CO
2
%
2
A display calibration adjustment must be entered to match the
display to the reference device. See next step.
Reference Device
X
Start
Begin Calibration
2 Hours Elapsed
(Minimum)

Table of Contents