54 | Page
MAINTENANCE
End of Procedure
CO
2
Calibration Continued
7
Compare the reference device reading with the incubator CO
2
display again.
If the reference device and the incubator display readings
are the same or the difference falls within the range of your
protocol, the incubator is now calibrated for CO
2
.
-OR-
See the next step if the readings fail to match or fall outside
of your protocol range.
Reference Device
8
If the difference still falls outside the acceptable range of your
protocol, repeat steps 3 – 7 up to two more times.
Reference Device
X
9
If the gas concentration readings of the incubator CO
2
display and the
reference device still fall outside your protocol after 3 calibration
attempts, contact your distributor or customer support for assistance.