2.6 Initial Filling or Rinsing of the Dosing unit
The dosing unit and the tubes are already mounted and ready-to-use. After the reagent bottle is connected, the
initial filling of the dosing unit can be performed. While performing the initial filling or washing program, an
adequately sized waste bin has to be placed below the titration tip.
On the main menu (fig. 9)
Fig. 9
Press <MODE> key and select Rinsing:
Fig. 10
Confirm the selection by pressing <OK>.
At this point you can select the number of rinsing cycles (Fig. 11). Initial filling requires a minimum of two rinsing
cycles. You can stop the rinsing operation (Fig. 12 and 13) at any time by pressing <STOP> and then resume
rinsing with <START>.
Fig. 11
The device fills first before the rinsing procedure starts: