EMC compatibility according to the Council Directive: 2004/108/EG;
applied harmonized standards: EN 61326-1:2012
Low-voltage directive according to the Council Directive 2006/95/EG
applied harmonized standards: EN 61010, part 1.
Country of origin: Germany, Made in Germany
The following dissolvents/titration reagents are allowed to be used:
All common titration solutions.
As reagent water and all non-aggressive non-organic and organic fluids are allowed. If using combustible
fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the chemical industry.
For fluids with higher viscosity ( 5 mm
2
/s), lower boiling point or affinity to outgas, the filling and dosage
speed can be adjusted.
Fluids with viscosity over 20mm
2
/s cannot be dosed.
Display: 3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Power supply: power supply 100-240 V; 50-60 Hz,
power input: 30 VA
Input: 12 Volt DC, 500 mA.
Use the Power supply TZ 1853, Type No.: FW 7362M/12 only!
RS-232-C Interface: RS-232-C interface separated, Daisy Chain function available.
Data bits: adjustable, 7 or 8 Bit (default: 8 Bit)
Stop bit: adjustable, 1 or 2 Bit (default: 1 Bit)
Start bit: static 1 Bit
Parity: adjustable: even / odd / none
Baud rate: adjustable: 1200, 2400, 4800, 9600, 19200 (Default 4800 baud)
Address: adjustable, (0 to 15, default: 01)
RS-232-1 for computer, input Daisy Chain
RS-232-2 devices of SI Analytics, titrator TitroLine
7750,
- Burettes TITRONIC
500, TITRONIC
110 plus, TITRONIC
universal/300,
- Balances of the types Mettler, Sartorius, Kern, Ohaus
(for more, please contact SI Analytics)
- Exit Daisy Chain
USB Interface: 1 x USB-type A and 1 x USB-type B
USB –Typ B for connecting a PC
USB –Typ A („Host“) for connecting:
- USB keyboard
- USB printer
- USB manual controller,
- USB data media e.g. USB stick
- USB Hub to connect more USB devices
Stirrer: Plug connection with integrated low-voltage supply (15 V ⎓) for the TM 50 Stirrer
incorporated in the bottom of the casing of the TITRONIC
®
300 piston burette